Abstract
Herein, gold nanoparticles (AuNP)-modified cortisol-imprinted (AuNP-MIP) plasmonic sensor was developed for signal amplification and real-time cortisol determination in both aqueous and complex solutions. Firstly, the sensor surfaces were modified with 3-(trimethoxylyl)propyl methacrylate and then pre-complex was prepared using the functional monomer N-methacryloyl-L-histidine methyl ester. The monomer solution was made ready for polymerization by adding 2-hydroxyethyl methacrylate to ethylene glycol dimethacrylate. In order to confirm the signal enhancing effect of AuNP, only cortisol-imprinted (MIP) plasmonic sensor was prepared without AuNP. To determine the selectivity efficiency of the imprinting process, the non-imprinted (AuNP-NIP) plasmonic sensor was also prepared without cortisol. The characterization studies of the sensors were performed with atomic force microscopy and contact angle measurements. The kinetic analysis of the AuNP-MIP plasmonic sensor exhibited a high correlation coefficient (R2 = 0.97) for a wide range (0.01–100 ppb) with a low detection limit (0.0087 ppb) for cortisol detection. Moreover, the high imprinting efficiency (k′ = 9.67) of the AuNP-MIP plasmonic sensor was determined by comparison with the AuNP-NIP plasmonic sensor. All kinetic results were validated and confirmed by HPLC.
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Cited by
22 articles.