Affiliation:
1. Key Laboratory of Photochemical Conversion and Optoelectronic Materials, Technical Institute of Physics and Chemistry, Chinese Academy of Sciences, Beijing 100190, China
2. University of Chinese Academy of Sciences, Beijing 100049, China
Abstract
The routine monitoring of bacterial populations is crucial for ensuring water quality and safeguarding public health. Thus, an electrochemical sensor based on a 1,10-phenanthroline-5,6-dione-modified electrode was developed and explored for the detection of E. coli. The modified electrode exhibited enhanced NADH oxidation ability at a low potential of 0.1 V, which effectively eliminated the interference from other redox compounds in bacteria. The sensitivity for NADH was 0.222 μA/μM, and the limit of detection was 0.0357 μM. Upon cell lysis, the intracellular NADH was released, and the concentration of E. coli was determined through establishing the relationship between the oxidation current signal and NADH concentration. The performance of the electrochemical sensor in the detection of NADH and E. coli suspensions was validated using the WST-8 colorimetric method. The blank recovery experiment in real water samples exhibited good accuracy, with recovery rates ranging from 89.12% to 93.26% and relative standard deviations of less than 10%. The proposed electrochemical sensor realized the detection of E. coli without the usage of biomarkers, which provides a promising approach for the broad-spectrum detection of microbial contents in complex water environments.
Funder
National Key Research and Development Program
Presidential Foundation of Technical Institute of Physics and Chemistry
Subject
Physical and Theoretical Chemistry,Analytical Chemistry