A Low-Cost Device for Measurement of Exhaled Breath for the Detection of Obstructive Lung Disease

Author:

Shahzad Adil AhmadORCID,Mushtaq Shafaq,Waris AsimORCID,Gilani Syed OmerORCID,Alnuwaiser Maha AbdallahORCID,Jameel MohammedORCID,Khan Niaz Bahadur

Abstract

Breath sensor technology can be used in medical diagnostics. This study aimed to build a device to measure the level of hydrogen sulfide, ammonia, acetone and alcohol in exhaled breath of patients as well as healthy individuals. The purpose was to determine the efficacy of these gases for detection of obstructive lung disease. This study was conducted on a total of 105 subjects, where 60 subjects were patients of obstructive lung disease and 45 subjects were healthy individuals. Patients were screened by means of the Pulmonary Function Test (PFT) by a pulmonologist. The gases present in the exhaled breath of all subjects were measured. The level of ammonia (32.29 ± 20.83 ppb), (68.83 ± 35.25 ppb), hydrogen sulfide (0.50 ± 0.26 ppm), (62.71 ± 22.20 ppb), and acetone (103.49 ± 35.01 ppb), (0.66 ± 0.31 ppm) in exhaled breath were significantly different (p < 0.05) between obstructive lung disease patients and healthy individuals, except alcohol, with a p-value greater than 0.05. Positive correlation was found between ammonia w.r.t Forced Expiratory Volume in 1 s (FEV1) (r = 0.74), Forced Vital Capacity (FVC) (r = 0.61) and Forced Expiratory Flow (FEF) (r = 0.63) and hydrogen sulfide w.r.t FEV1 (r = 0.54), FVC (r = 0.41) and FEF (r = 0.37). Whereas, weak correlation was found for acetone and alcohol w.r.t FEV1, FVC and PEF. Therefore, the level of ammonia and hydrogen sulfide are useful breath markers for detection of obstructive lung disease.

Publisher

MDPI AG

Subject

Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)

Reference24 articles.

1. Breath Analysis for Detecting Diseases on Respiratory, Metabolic and Digestive System

2. A selective nanosensor device for exhaled breath analysis

3. The versatile use of exhaled volatile organic compounds in human health and disease

4. MobSpiro: Mobile based spirometry for detecting COPD;Zubaydi;Proceedings of the 2017 IEEE 7th Annual Computing and Communication Workshop and Conference (CCWC),2017

5. Guidelines for the Diagnosis and Management of Asthma,1991