Abstract
Abstract
Analyzing exhaled breath samples, especially using a highly sensitive method such as MCC/IMS (multi-capillary column/ion mobility spectrometry), may also detect analytes that are derived from exogenous production. In this regard, there is a discussion about the optimal interpretation of exhaled breath, either by considering volatile organic compounds (VOCs) only in exhaled breath or by additionally considering the composition of room air and calculating the alveolar gradients. However, there are no data on whether the composition and concentration of VOCs in room air are identical to those in truly inhaled air directly before analyzing the exhaled breath. The current study aimed to determine whether the VOCs in room air, which are usually used for the calculation of alveolar gradients, are identical to the VOCs in truly inhaled air. For the measurement of inhaled air and room air, two IMS, each coupled with an MCC that provided a pre-separation of the VOCs, were used in parallel. One device was used for sampling room air and the other for sampling inhaled air. Each device was coupled with a newly invented system that cleaned room air and provided a clean carrier gas, whereas formerly synthetic air had to be used as a carrier gas. In this pilot study, a healthy volunteer underwent three subsequent runs of sampling of inhaled air and simultaneous sampling and analysis of room air. Three of the selected 11 peaks (P4—unknown, P5—1-Butanol, and P9—Furan, 2-methyl-) had significantly higher intensities during inspiration than in room air, and four peaks (P1—1-Propanamine, N-(phenylmethylene), P2—2-Nonanone, P3—Benzene, 1,2,4-trimethyl-, and P11—Acetyl valeryl) had higher intensities in room air. Furthermore, four peaks (P6—Benzaldehyde, P7—Pentane, 2-methyl-, P8—Acetone, and P10—2-Propanamine) showed inconsistent differences in peak intensities between inhaled air and room air. To the best of our knowledge, this is the first study to compare simultaneous sampling of room air and inhaled air using MCC/IMS. The simultaneous measurement of inhaled air and room air showed that using room air for the calculation of alveolar gradients in breath analysis resulted in different alveolar gradient values than those obtained by measuring truly inhaled air.
Subject
Pulmonary and Respiratory Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献