Abstract
A notable diagnostic for the detection of hemolytic diseases is bilirubin, a by-product of haemoglobin breakdown. The concentration of bilirubin ranges from 0.3 to 1.9 mg in 100 mL of blood. Low blood bilirubin levels are associated with a greater risk of coronary heart disease and anaemia. Hyperbilirubinemia results from a serum bilirubin level of more than 2.5 mg/100 mL. Therefore, it is very crucial to check the serum bilirubin level. Analytical equipment for point-of-care testing must be portable, small, and affordable. A unique method is used to detect bilirubin selectively using paper-based screen-printed carbon electrodes that were covalently linked with nanoparticles, that serves as a key biomarker for jaundice. In order to create an electrochemical biosensor, bilirubin oxidase was immobilised on electrodes modified with AgNPs. The morphology of Ag nanoparticles in terms of size and shape was determined using both UV- Vis Spectroscopy and transmission electron microscopy (TEM). The biosensor’s analytical response was assessed using potentiostat (Cyclic voltammetry (CV) and linear sweep voltammetry (LSV)). The developed paper-based sensor provided optimum feedback and a broad linear range of 1 to 9 µg/mL for bilirubin, with a lower LOD of 1 µg/mL. Through tests of bilirubin in artificial blood serum, the viability is confirmed. The method that is being used makes it possible to create and use an inexpensive, miniature electrochemical sensor.
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献