Abstract
The measurement and analysis of the arterial pulse wave provides information about the state of vascular health. When measuring blood pressure according to Riva-Rocci, the systolic and diastolic blood pressure is measured non-invasively with an inflatable pressure cuff on the upper arm. Today’s blood pressure monitors analyze the pulse wave in reference to the rising or falling cuff pressure. With the help of additional pulse wave analysis, one can determine the pulse rate and the heart rate variability. In this paper, we investigated the concept, the construction, and the limitations of ultrawideband (UWB) radar and continuous wave (CW) radar, which provide continuous and non-invasive pulse wave measurements. We integrated the sensors into a complete measurement system. We measured the pulse wave of the cuff pressure, the radar sensor (both UWB and CW), the optical sensor, and ultrasonic Doppler as a reference. We discussed the results and the sensor characteristics. The main conclusion was that the resolution of the pulse radar was too low, even with a maximum bandwidth of 10 GHz, to measure pulse waves reliably. The continuous wave radar provides promising results for a phantom if adjusted properly with phase shifts and frequency. In the future, we intend to develop a CW radar solution with frequency adaption.
Funder
Deutsche Forschungsgemeinschaft
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献