Monitoring System for Operating Variables in Incubators in the Neonatology Service of a Highly Complex Hospital through the Internet of Things (IoT)
Author:
Aya-Parra Pedro Antonio1, Rodriguez-Orjuela Andres Jacob12, Rodriguez Torres Viviana3, Cordoba Hernandez Nidia Patricia4, Martinez Castellanos Natalia4, Sarmiento-Rojas Jefferson1ORCID
Affiliation:
1. Biomedical Engineering Program, School of Medicine and Health Sciences, Universidad del Rosario, Bogota 111711, Colombia 2. Biomedical Engineering Program, Engineering Faculty, Escuela Colombiana de Ingeniería Julio Garavito, Bogota 110111, Colombia 3. Departamento de Neonatología, Hospital Universitario Mayor—Méderi, Bogota 110111, Colombia 4. Departamento de Ingeniería Biomédica, Hospital Universitario Mayor—Méderi, Bogota 110111, Colombia
Abstract
Background: Around 15 million premature babies are born annually, requiring specialized care. Incubators are vital for maintaining their body temperature, which is crucial for their well-being. Ensuring optimal conditions in incubators, including constant temperature, oxygen control, and comfort, is essential for improving the care and survival rates of these infants. Methods: To address this, an IoT-based monitoring system was developed in a hospital setting. The system comprised hardware components such as sensors and a microcontroller, along with software components including a database and a web application. The microcontroller collected data from the sensors, which was then transmitted to a broker via WiFi using the MQTT protocol. The broker validated and stored the data in the database, while the web application provided real-time access, alerts, and event recording. Results: Two certified devices were created, employing high quality components. The system was successfully implemented and tested in both the biomedical engineering laboratory and the neonatology service of the hospital. The results of the pilot test supported the concept of IoT-based technology, demonstrating satisfactory responses in temperature, humidity, and sound variables within the incubators. Conclusions: The monitoring system facilitated efficient record traceability, allowing access to data over various timeframes. It also captured event records (alerts) related to variable problems, providing information on duration, date, hour, and minutes. Overall, the system offered valuable insights and enhanced monitoring capabilities for neonatal care.
Funder
Hospital Universitario Mayor
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Reference28 articles.
1. (2023, May 25). Nacimientos Prematuros. Available online: https://www.who.int/es/news-room/fact-sheets/detail/preterm-birth. 2. Global, regional, and national causes of under-5 mortality in 2000–2015: An updated systematic analysis with implications for the Sustainable Development Goals;Liu;Lancet,2016 3. Cuando su Bebé Está en la Unidad de Cuidados Intensivos Neonatales (UCIN) (2023, May 25). Kidshealth.org. Available online: https://kidshealth.org/es/parents/nicu-caring.html. 4. Fernanda, L., Revisado, E., and Lic, L.G. (2023, May 25). El ABC del Cuidado de Enfermería en los Bebés Prematuros Extremos–Segunda Parte–. Available online: https://www.fundasamin.org.ar/archivos/EL%20ABC%20DEL%20CUIDADO%20DE%20ENFERMERIA.pdf. 5. Pardell, X. (2023, May 25). Incubadora Neonatal—Apuntes de Electromedicina Xavier Pardell. Available online: https://www.pardell.es/incubadora-neonatal.html.