Evaluation of the Dimensional Accuracy of Robot-Guided Laser Osteotomy in Reconstruction with Patient-Specific Implants—An Accuracy Study of Digital High-Tech Procedures

Author:

Msallem Bilal12ORCID,Veronesi Lara23,Beyer Michel23,Halbeisen Florian S.4ORCID,Maintz Michaela25,Franke Adrian6ORCID,Korn Paula6,Dragu Adrian1ORCID,Thieringer Florian M.23ORCID

Affiliation:

1. University Center for Orthopedics, Trauma and Plastic Surgery, Medical Faculty and University Hospital Carl Gustav Carus, Technische Universität Dresden, DE-01307 Dresden, Germany

2. Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, CH-4123 Allschwil, Switzerland

3. Clinic of Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, CH-4031 Basel, Switzerland

4. Surgical Outcome Research Center, Department of Clinical Research, University of Basel c/o University Hospital of Basel, CH-4001 Basel, Switzerland

5. Institute for Medical Engineering and Medical Informatics, University of Applied Sciences and Arts Northwestern Switzerland, CH-4132 Muttenz, Switzerland

6. Department of Oral and Maxillofacial Surgery, Medical Faculty and University Hospital Carl Gustav Carus, Technische Universität Dresden, DE-01307 Dresden, Germany

Abstract

Background/Objective: With the rapid advancement in surgical technologies, new workflows for mandibular reconstruction are constantly being evaluated. Cutting guides are extensively employed for defining osteotomy planes but are prone to errors during fabrication and positioning. A virtually defined osteotomy plane and drilling holes in robotic surgery minimize potential sources of error and yield highly accurate outcomes. Methods: Ten mandibular replicas were evaluated after cutting-guided saw osteotomy and robot-guided laser osteotomy following reconstruction with patient-specific implants. The descriptive data analysis summarizes the mean, standard deviation (SD), median, minimum, maximum, and root mean square (RMS) values of the surface comparison for 3D printed models regarding trueness and precision. Results: The saw group had a median trueness RMS value of 2.0 mm (SD ± 1.7) and a precision of 1.6 mm (SD ± 1.4). The laser group had a median trueness RMS value of 1.2 mm (SD ± 1.1) and an equal precision of 1.6 mm (SD ± 1.4). These results indicate that robot-guided laser osteotomies have a comparable accuracy to cutting-guided saw osteotomies, even though there was a lack of statistical significance. Conclusions: Despite the limited sample size, this digital high-tech procedure has been shown to be potentially equivalent to the conventional osteotomy method. Robotic surgery and laser osteotomy offers enormous advantages, as they enable the seamless integration of precise virtual preoperative planning and exact execution in the human body, eliminating the need for surgical guides in the future.