Abstract
This work demonstrated a sensitive, selective, and simple liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for quantitation of sumatriptan in human plasma samples. Terazosin was used as an internal standard to minimize the variability during sample processing and detection. Sample cleanup prior to chromatographic analysis was accomplished by liquid-liquid extraction (LLE) with tert-butyl methyl ether (t-BME). The separation was performed on a reversed-phase Symmetry® C18 column (150 × 4.6 mm i.d., 5 µm) under a gradient mode, using a 0.2% formic acid aqueous solution and acetonitrile at a flow rate of 0.5 mL/min. Sumatriptan (m/z 296.26→251.05) and terazosin (m/z 388.10→290.25) were quantified using a triple quadrupole mass spectrometer, operating in the multiple reaction monitoring (MRM) under the positive ion mode. The method was fully validated following US-FDA and EMA guidelines. The LC-MS/MS assay had a calibration range of 0.5–50.0 ng/mL. The assay was precise and accurate with a between-run precision of <9.51%, and between-run accuracy between −7.27 to 8.30%. The developed method was subsequently applied in the determination of plasma concentration-time profile of a sumatriptan 50-mg tablet following oral administration in healthy volunteers.
Subject
Drug Discovery,Pharmaceutical Science,Molecular Medicine
Cited by
10 articles.
订阅此论文施引文献