Gas Adsorption Response of Piezoelectrically Driven Microcantilever Beam Gas Sensors: Analytical, Numerical, and Experimental Characterizations

Author:

Nsubuga Lawrence1ORCID,Duggen Lars2,Marcondes Tatiana Lisboa3,Høegh Simon3,Lofink Fabian4,Meyer Jana4,Rubahn Horst-Günter1ORCID,de Oliveira Hansen Roana1ORCID

Affiliation:

1. SDU NanoSYD, Mads Clausen Institute, University of Southern Denmark, Alsion 2, 6400 Sønderborg, Denmark

2. SDU Mechatronics, Department of Mechanical and Electrical Engineering, University of Southern Denmark, Alsion 2, 6400 Sønderborg, Denmark

3. AmiNIC ApS, Jernbanegade 75, 5500 Middlefart, Denmark

4. Fraunhofer Institute for Silicon Technology, Fraunhoferstraße 1, 25524 Itzehoe, Germany

Abstract

This work presents an approach for the estimation of the adsorbed mass of 1,5-diaminopentane (cadaverine) on a functionalized piezoelectrically driven microcantilever (PD-MC) sensor, using a polynomial developed from the characterization of the resonance frequency response to the known added mass. This work supplements the previous studies we carried out on the development of an electronic nose for the measurement of cadaverine in meat and fish, as a determinant of its freshness. An analytical transverse vibration analysis of a chosen microcantilever beam with given dimensions and desired resonance frequency (>10 kHz) was conducted. Since the beam is considered stepped with both geometrical and material non-uniformity, a modal solution for stepped beams, extendable to clamped-free beams of any shape and structure, is derived and used for free and forced vibration analyses of the beam. The forced vibration analysis is then used for transformation to an equivalent electrical model, to address the fact that the microcantilever is both electronically actuated and read. An analytical resonance frequency response to the mass added is obtained by adding simulated masses to the free end of the beam. Experimental verification of the resonance frequency response is carried out, by applying known masses to the microcantilever while measuring the resonance frequency response using an impedance analyzer. The obtained response is then transformed into a resonance frequency to the added mass response polynomial using a polynomial fit. The resulting polynomial is then verified for performance using different masses of cantilever functionalization solution. The functionalized cantilever is then exposed to different concentrations of cadaverine while measuring the resonance frequency and mass of cadaverine adsorbed estimated using the previously obtained polynomial. The result is that there is the possibility of using this approach to estimate the mass of cadaverine gas adsorbed on a functionalized microcantilever, but the effectiveness of this approach is highly dependent on the known masses used for the development of the response polynomial model.

Funder

Danmarks Frie Forskningsfond

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Reference65 articles.

1. Microcantilever-based sensors;Chaudhary;Def. Sci. J.,2009

2. Microcantilever-based biosensor for detection of various biomolecules;Yoo;Jpn. J. Appl. Phys.,2006

3. A review of microcantilevers for sensing applications;Vashist;J. Nanotechnol.,2007

4. Towards non-linear modeling of molecular interactions arising from adsorbed biological species on the microcantilever surface;Afshari;Int. J. -Non-Linear Mech.,2007

5. Microcantilever sensors;Thundat;Microscale Thermophys. Eng.,1997

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献