Technical note: Realization and uncertainty analysis for an adjustable 3D structured breast phantom in digital breast tomosynthesis

Author:

Salomon Elisabeth1,Unger Ewald1,Homolka Peter1,Cockmartin Lesley2,Petrov Dimitar2,Semturs Friedrich1,Songsaeng Chatsuda13,Panagiotis Kapetas4,Vancoillie Liesbeth2,Figl Michael13,Sommer Alexander5,Bosmans Hilde2,Hummel Johann1

Affiliation:

1. Center for Medical Physics and Biomedical Engineering Medical University of Vienna, Spitalgasse Vienna Austria

2. Department of Radiology UZ Gasthuisberg Leuven Belgium

3. Christian Doppler Laboratory for Mathematical Modelling and Simulation of Next‐Generation Medical Ultrasound Devices Vienna Austria

4. Department of Biomedical Imaging and Image‐guided Therapy Medical University of Vienna and General Hospital, Spitalgasse Vienna Austria

5. Clinic for Radiology and Reference Center for Mammography Münster University of Münster and University Hospital Münster Münster Germany

Abstract

AbstractBackgroundProjection imaging phantoms are often optimized for 2‐dimensional image characteristics in homogeneous backgrounds. Therefore, evaluation of image quality in tomosynthesis (DBT) lacks accepted and established phantoms.PurposeWe describe a 3D breast phantom with a structured, variable background. The phantom is an adaptable and advanced version of the L1 phantom by Cockmartin et al. Phantom design and its use for quality assurance measurements for DBT devices are described. Four phantoms were compared to assess the objectivity.MethodsThe container size was increased to a diameter of 24 cm and a total height of 53.5 mm. Spiculated masses were replaced by five additional non‐spiculated masses for higher granularity in threshold diameter resolution. These patterns are adjustable to the imaging device. The masses were printed in one session with a base layer using two‐component 3D printing. New materials compared to the L1 phantom improved the attenuation difference between the lesion models and the background. Four phantoms were built and intra‐human observer, inter‐human observer and inter‐phantom variations were determined. The latter assess the reproducibility of the phantom production. Coefficients of variance (V) were calculated for all three variations.ResultsThe difference of the attenuation coefficients between the lesion models and the background was 0.20 cm−1 (with W/Al at 32 kV, equivalent to 19–20 keV effective energy) compared to 0.21 cm−1 for 50/50 glandular/adipose breast tissue and cancerous lesions. PMMA equivalent thickness of the phantom was 47.0 mm for the Siemens Mammomat Revelation. For the masses, the for the intra‐observer variation was 0.248, the averaged inter‐observer variation, was 0.383. for phantom variance was 0.321. For the micro‐calcifications, was 0.0429, 0.0731 and 0.0759.ConclusionsPosition, orientation and shape of the masses are reproducible and attenuation differences appropriate. The phantom presented proved to be a candidate test object for quality control.

Funder

Christian Doppler Forschungsgesellschaft

Publisher

Wiley

Subject

General Medicine

Reference31 articles.

1. LindellB DunsterH ValentinJ.International Commission on Radiological Protection: History Policies and Procedures.https://euref.org/european‐guidelines/1998.

2. vanEngenR et al.European guidelines for quality assurance in breast cancer screening and diagnosis.https://euref.org/european‐guidelines (2006).

3. Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systems

4. Radiation risk of breast screening in England with digital mammography

5. Quality control for digital mammography in the ACRIN DMIST trial: Part I

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献