Affiliation:
1. Universidade Federal da Grande Dourados Rodovia Dourados-Itahum, km 12. Dourados, MS CEP 79804-970 Brazil
2. National Institute for Alternative Technologies of Detection Toxicological Evaluation and Removal of Micropollutants and Radioactives Institute of Chemistry UNESP Araraquara, SP CEP 14800–900 Brazil
Abstract
AbstractThe combination of 3D‐printing technology and microfluidics has garnered widespread interest for creating new designs and has opened a large window for creativity. However, some laboratories do not have the necessary technology to explore this potential advantage. In this paper, we demonstrate how to explore target technologies to fabricate a single version of a polydimethylsiloxane (PDMS)‐based electroanalytical platform using a low‐cost protocol and disposable materials. This protocol is cost‐effective in terms of equipment and materials, allowing for the creation of a circular channel, a T‐junction system, and two designs of electrochemical detectors customized onto a single platform. The interesting new device has some important characteristics, such as fabrication following green and sustainable protocols. Additionally, it can address the limitations associated with flow‐based analysis, for which the classical three‐electrode configuration is inadequate. It has also demonstrated the feasibility of using the device in analytical applications for the determination of salicylic acid in commercial dermocosmetic samples, including aqueous samples.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献