Haemostasis after partial hepatectomy using argon beam coagulation

Author:

Postema R R1,Plaisiert P W1,Ten Kate F J W2,Terpstra O T314

Affiliation:

1. Department of General Surgery, University Hospital Dijkzigt, Rotterdam, The Netherlands

2. Department of Pathology, University Hospital Dijkzigt, Rotterdam, The Netherlands

3. Laboratory for Experimental Surgery, Erasmus University, The Netherlands

4. Department of General Surgery, University Hospital, Leiden, The Netherlands

Abstract

Abstract The argon beam coagulator is a new device for haemostasis during and after surgery on parenchymatous organs. No data are available on its efficacy and tissue effect following hepatic resection. Blood loss, the time needed to achieve adequate haemostasis and histological findings after liver resection were assessed in 12 pigs using argon beam coagulation or suture ligation only, the mattress suture technique and tissue glue application. The treatment was randomly assigned to each of the four liver lobes in each pig. Median blood loss following argon beam coagulation was 13 (range 2–47) ml and after simple suture ligation 55 (range 2–260) ml (P < 0·02). The median time needed for adequate haemostasis following argon beam coagulation was 3 (range 2–7) min versus 14 (range 2–48) min in the control group (P < 0·005). There was no difference between argon beam coagulation and tissue glue, which were both superior to the use of mattress sutures. Argon beam coagulation resulted in less tissue damage than tissue glue or mattress suturing. The argon beam coagulator is an efficient device for achieving haemostasis following partial hepatectomy in the pig. It causes only a moderate tissue reaction.

Publisher

Oxford University Press (OUP)

Subject

Surgery

Reference6 articles.

1. Electrosurgical deblking of ovarian cancer: a new technique using the argon beam coagulator;Brand;Gynecol Oncol,1990

2. A significant new contribution to radical head and neck surgery. The argon beam coagulator as an effective means of limiting blood loss;Ward;Arch Otolaryngol Head Neck Surg,1989

3. A controlled study of the argon beam coagulator for partial nephrectomy;Hernandez;J Urol,1990

4. Liver transection: a controlled study of four different techniques in pigs;Ottow;Surgery,1985

5. Repair of the resected liver stump: an experimental study;Tzuzuki;Surgery,1972