The influence of implant design and limb alignment on in vivo wear rates of fixed‐bearing and rotating‐platform knee implant retrievals

Author:

Dreyer Michael J.12ORCID,Weisse Bernhard2,Contreras Raggio José Ignacio23ORCID,Zboray Robert4,Taylor William R.1ORCID,Preiss Stefan5,Horn Nils5

Affiliation:

1. Laboratory for Movement Biomechanics, Institute for Biomechanics ETH Zürich Zürich Switzerland

2. Laboratory for Mechanical Systems Engineering Empa Dübendorf Switzerland

3. Facultad de Ingeniería y Ciencias Universidad Adolfo Ibáñez Viña del Mar Chile

4. Center for X‐ray Analytics Empa Dübendorf Switzerland

5. Department of Lower Extremities Schulthess Clinic Zürich Switzerland

Abstract

AbstractAnalysis of polyethylene (PE) wear in knee implants is crucial for understanding the factors leading to revision in total knee arthroplasty. Importantly, current experimental and computational methods for predicting insert wear can only be validated against true in vivo measurements from retrievals. This study quantitatively investigated in vivo PE wear rates in fixed‐bearing (FB) (n = 21) and rotating‐platform (n = 53) implant retrievals. 3D surface geometry of the retrievals was measured using a structured light scanner. Then, a reference surface that included the deformation, but not the wear that the retrievals had experienced in vivo, was constructed using a fully automatic surface reconstruction algorithm. Finally, wear volume was calculated from the deviation between the worn and reconstructed surfaces. The measurement and analysis techniques were validated and the algorithm was found to produce errors of only 0.2% relative to the component volumes. In addition to quantifying cohort‐level wear rates, the effect of mechanical axis limb alignment on mediolateral wear distribution was examined for a subset of the retrievals (n = 14 + 26). Our results show that FB implants produce significantly (p = 0.04) higher topside wear rates (24.6 ± 10.1 mm3/year) than rotating‐platform implants (15.3 ± 8.0 mm3/year). This effect was larger than that of limb alignment, which had a smaller and nonsignificant influence on overall wear rates (+4.5 ± 11.6 mm3/year, p = 0.43). However, increased varus alignment was associated significantly with greater medial compartment wear in both the FB and rotating‐platform designs (+1.7 ± 1.3%/° and +1.8 ± 1.6%/°). Our findings emphasize the importance of implant design and limb alignment on wear outcomes, providing reference data for improving implant performance and longevity.

Publisher

Wiley

Subject

Orthopedics and Sports Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3