Comparison between a handheld ultrasound device and a traditional ultrasound for performing transcranial sonography in patients with Parkinson's disease

Author:

Paes Maria A. S.1ORCID,Nicaretta Denise H.2ORCID,Alvarenga Regina M. P.2ORCID,Rosso Ana L. Z.3ORCID,Brisson Rodrigo T.3ORCID,Fernandes Rita C.L.3ORCID

Affiliation:

1. Departamento de Radiologia Hospital Universitário Gaffrée e Guinle Rio de Janeiro Brazil

2. Departamento de Neurologia Hospital Universitário Gaffrée e Guinle Rio de Janeiro Brazil

3. Faculdade de Medicina Hospital Universitário Clementino Fraga Filho Rio de Janeiro Brazil

Abstract

AbstractObjectiveThe aim of this study is to compare a portable ultrasound (US) device and a traditional US for performing transcranial ultrasonography (CCT) in patients with Parkinson's disease (PD).MethodsThis is a cross‐sectional, observational, and analytical study. The study recruited a total of 129 individuals from two public hospitals in the city of Rio de Janeiro in a prospective and non‐randomized manner between September 2019 and July 2021 as follows: group A with 31 patients with PD, group B with 65 patients with PD, and group C with 64 healthy individuals. Group A was used to collect data to establish the agreement analysis of the TCS measurements between the two devices. Groups B and C provided data for constructing the receiver operating characteristic curve for the handheld US. The subjects underwent the assessment of the transtemporal bone window (TW) quality, the mesencephalon area, the size of the third ventricle, and the substantia nigra (SN) hyperechogenicity area.ResultsThere was a good agreement between the methods regarding the quality of the TW–Kappa concordance coefficient of 100% for the right TW and 83% for the left, the midbrain area—intraclass correlation coefficient (ICC) of 69%, the SN area ICC = 90% for the right SN and 93% for the left and the size of the third ventricle ICC = 96%. The cutoff point for the SN echogenic area in the handheld US was 0.20 cm2.ConclusionsThe handheld US is a viable imaging method for performing TCS because it shows good agreement with the measurements performed with traditional equipment, and the measurement of SN echogenic area for PD diagnosis presents good sensitivity and specificity.

Publisher

Wiley

Subject

Behavioral Neuroscience

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3