An in vitro study evaluating the efficacy of a novel mount with torque control designed to tighten Osstell® transducers

Author:

Naughton David1,Donnelly‐Swift Erica23,Polyzois Ioannis1ORCID

Affiliation:

1. Department of Restorative Dentistry and Periodontology Dublin Dental University Hospital, Trinity College Dublin Ireland

2. Department of Public Health and Primary Care Institute of Population Health, Trinity College Dublin Dublin Ireland

3. Dublin Dental University Hospital, Trinity College Dublin Dublin Ireland

Abstract

AbstractObjectivesTo evaluate the efficacy of a novel mount with torque control for tightening of Osstell® transducers and to determine the reliability of recorded ISQ measurements from implants placed in various bone densities.Material and MethodsFifty‐six implants, comprising seven different implant types, were placed in eight polyurethane blocks representing D1, D2, D3, and D4 bone densities. Resonance frequency analysis (RFA) transducers were attached to each implant in four different ways: (a) hand tightening, (b) hand tightening with a SmartPeg Mount™, (c) hand tightening using the novel mount with torque control (SafeMount) and (d) tightening to 6 Ncm with a calibrated torque device. ISQ measurements were taken and a second operator repeated the measurements. Intraclass correlation coefficient (ICC) was calculated to assess the reliability of the measurements and linear mixed effects regression was employed to determine the effect explanatory variables had on ISQ values.ResultsThere was a statistically significant difference in ISQ values obtained by hand tightening transducers compared to the calibrated torque device p < .001, 95%(−2.89, −1.21) but not between any other tightening methods. There was excellent agreement between the two RFA devices (ICC 0.986) and between buccal and mesial measurements (ICC 0.977). For all transducer tightening methods there was excellent inter‐operator agreement in D1 and D2 (ICC > 0.8) but very poor agreement in D4 (ICC < 0.24). Bone density accounted for 36% of the variation in ISQ values, the implant for 11% and the operator for 6%.ConclusionsSafeMount, did not significantly improve the reliability of the RFA measurements when compared to the standard mount, but calibrated torque devices seem to have benefits when compared to tightening the transducers by hand. Results also indicate that the ISQ values should be interpreted with caution when measuring implant stability in poor quality bone regardless of the implant geometry.

Publisher

Wiley

Subject

General Dentistry

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3