Effect of the Field of View Size on CBCT Artifacts Caused by the Presence of Metal Objects in the Exomass

Author:

Safi Yaser1ORCID,Ghazizadeh Ahsaie Mitra1ORCID,Jafarian Amiri Maede1ORCID

Affiliation:

1. Department of Oral and Maxillofacial Radiology, School of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Abstract

Background and Aim. This study aimed to assess the effect of the size of the field of view (FOV) on cone-beam computed tomography (CBCT) artifacts caused by metal objects in the exomass. Materials and Methods. In this in vitro experimental study, titanium implants, teeth with cobalt-chromium (Co-Cr) intracanal posts, and teeth with mesio-occluso-distal (MOD) amalgam restorations were placed in an empty socket of the extracted third molar of a human mandible. These metallic materials were differently arranged in the exomass (zone outside of the FOV). A polypropylene tube containing dipotassium phosphate was placed in the empty socket of the right canine tooth in a dry human mandible. CBCT scans were taken with a NewTom VGI (Verona, Italy) scanner using a 6 × 6 cm and an 8 × 8 cm FOV. The histogram tool of OnDemand software (Cybermed, Seoul, Korea) was used to select circles with a 1.5 mm diameter as the (ROI) at the center of the homogenous solution of dipotassium phosphate tube on the axial plane. The mean gray value (GV) and its standard deviation (SD) in the region of interest (ROI) were calculated ( P > 0.05 ). The data were analyzed by SPSS 26. Results. The reduction in the size of the FOV significantly decreased the mean GV ( P < 0.001 ). Metal objects in the exomass significantly decreased the mean GV ( P < 0.001 ), and minimum mean GV and maximum SD were recorded for amalgam, followed by Co-Cr intracanal posts, and titanium implants. The unilateral presence of a metal object was associated with a higher mean GV and lower SD ( P < 0.001 ). Conclusion. Using a smaller FOV increases the size of the exomass, which may negatively affect the image quality. Metal objects in the exomass decrease the GV of CBCT scans and adversely affect the image quality.

Publisher

Hindawi Limited

Subject

General Dentistry

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3