Affiliation:
1. Department of Urology, Seoul St. Mary’s Hospital, 222 Banpo-daero, Seocho-gu, Seoul 137701, Republic of Korea
2. Department of Mechanical Engineering, Sogang University, Seoul, Republic of Korea
Abstract
Polydimethylsiloxane (PDMS) is used in implantable medical devices; however, PDMS is not a completely biocompatible material for electronic medical devices in the bladder. To identify novel biocompatible materials for intravesical implanted medical devices, we evaluated the biocompatibility of polymethyl methacrylate (PMMA) by analyzing changes in the levels of macrophages, macrophage migratory inhibitory factor (MIF), and inflammatory cytokines in the bladder. A ball-shaped metal coated with PMMA or PDMS was implanted into the bladders of rats, and after intravesical implantation, the inflammatory changes induced by the foreign body reaction were evaluated. In the early period after implantation, increased macrophage activity and MIF in the urothelium of the bladder were observed. However, significantly decreased macrophage activity and MIF in the bladder were observed after implantation with PMMA- or PDMS-coated metal in the later period. In addition, significantly decreased inflammatory cytokines such as IL-1β, IL-6, and TNF-αwere observed with time. Based on these results, we suggest that MIF plays a role in the foreign body reaction and in the biocompatible packaging with PMMA for the implanted medical devices in the bladder.
Funder
Ministry of Health and Welfare
Subject
General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine
Cited by
57 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献