Determination of Electrolytes in Critical Illness Patients at Different pH Ranges: Whom Shall We Believe, the Blood Gas Analysis or the Laboratory Autoanalyzer?

Author:

Hohmann Christopher1ORCID,Pfister Roman1,Kuhr Kathrin2,Merkle Julia3,Hinzmann Julian1,Michels Guido1

Affiliation:

1. Department III of Internal Medicine, Heart Center, University Hospital of Cologne, Cologne, Germany

2. Institute of Medical Statistics and Computational Biology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Cologne, Germany

3. Department of Cardiothoracic Surgery, Heart Center, University Hospital of Cologne, Cologne, Germany

Abstract

Introduction. The determination of the electrolytes sodium and potassium is essential in critical care. In daily clinical practice, both the blood gas analyzer (ABG) and the laboratory autoanalyzer (AA) are generally applied. However, there is still uncertainty regarding the convergence of the prementioned assays, and data about the comparability dependent on the pH value are still lacking.Materials and Methods. One hundred samples from intensive care unit patients with a range in pH values between 7.20 and 7.49 were evaluated in this retrospective cohort study. All patients suffered an infarct-related cardiogenic shock and were intubated and not under therapeutical hypothermia at the time of blood collection. We used scatter plots to compare different distributions of sodium and potassium values between the methods. Comparability of the analyses was assessed using the Bland–Altmann approach, and intraclass correlations (ICC) as estimates of interrater reliability were calculated.Results. The mean potassium level measured on ABG was 4.33 mmol/L (SD 0.48 mmol/L), and the value obtained using the AA was 4.40 mmol/L (SD 0.55 mmol/L). A Bland–Altman comparison for total potassium measurements revealed that the limits of agreement were small (−0.241 to 0.391 mmol/L). Total ICC displayed a very good correlation of 0.949. For sodium, we found average values of 140 mmol/L (SD 5.20 mmol/L) in the AA and 140 mmol/L (SD 5.80 mmol/L) in the ABG assessment. Contrarily, the Bland–Altman comparison for sodium displayed that the 95% limits of agreement were very wide (−5.99 to 6.59 mmol/L) for total measurements as well as in every pH subgroup. Total ICC only reached a value of 0.830.Conclusion. Data from our single-center study indicate that urgent and vital decisions based on potassium measurements can be made by trusting the value obtained on the ABG machine irrespective of pH values.

Publisher

Hindawi Limited

Subject

Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3