Comparison of Point-of-Care Versus  Central Laboratory Measurement of Electrolyte Concentrations on Calculations of the Anion Gap and the Strong Ion Difference

Author:

Morimatsu Hiroshi1,Rocktäschel Jens1,Bellomo Rinaldo2,Uchino Shigehiko1,Goldsmith Donna3,Gutteridge Geoffrey4

Affiliation:

1. Research Fellow.

2. Director of Research.

3. Research Coordinator.

4. Director, Department of Intensive Care, Austin, and Repatriation Medical Centre.

Abstract

Background Clinicians calculate the anion gap (AG) and the strong ion difference (SID) to make acid-base diagnoses. The technology used is assumed to have limited impact. The authors hypothesized that different measurement technologies markedly affect AG and SID values. Methods SID and AG were calculated using values from the point-of-care blood gas and electrolyte analyzer and the central hospital laboratory automated blood biochemistry analyzer. Simultaneously measured plasma sodium, potassium, and chloride concentrations were also compared. Results Mean values for central laboratory and point-of-care plasma sodium concentration were significantly different (140.4 +/- 5.6 vs. 138.3 +/- 5.9 mm; P < 0.0001), as were those for plasma chloride concentration (102.4 +/- 6.5 vs. 103.4 +/- 6.0 mm; P < 0.0001) but not potassium. Mean AG values calculated with the two different measurement techniques differed significantly (17.6 +/- 6.2 mEq/l for central laboratory vs. 14.5 +/- 6.0 mEq/l for point-of-care blood gas analyzer; P < 0.0001). Using the Stewart-Figge methodology, SID values also differed significantly (43.7 +/- 4.8 vs. 40.7 +/- 5.6 mEq/l; P < 0.0001), with mean difference of 3.1 mEq/l (95% limits of agreement, -3.4, 9.5 mEq/l). For 83 patients (27.6%), differences in AG values were as high as 5 mEq/l or more, and for 46% of patients whose AG value was outside the reference range with one technology, a value within normal limits was recorded with the other. Conclusions Results with two different measurement technologies differed significantly for plasma sodium and chloride concentrations. These differences significantly affected the calculated AG and SID values and might lead clinicians to different assessments of acid-base and electrolyte status.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference24 articles.

Cited by 95 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3