Abstract
Water content in jet fuels is detected by thermometric titration (TMT), and the optimal detected system is 2,2-dimethoxypropane as titrant, cyclohexane and isopropanol as titration solvents, and methanesulfonic acid as catalyst in this method. The amounts of oil, concentration and delivery rate of titrant, volumes, and the reliability and accuracy of thermometric titration were emphasized. The results show that the accuracy, validity, and reliability of TMT are excellent by different indicated spiked water contents. The obtained results between TMT and Karl Fischer titration have been proven to be in accord. But, the duration of titration merely spends 3–5 min in the whole process, greatly shortening the detected time. Therefore, rapid and accurate determination of trace water in a jet fuel can be realized by TMT.
Funder
Tribology Science Fund of State Key Laboratory of Tribology
Subject
Spectroscopy,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献