Bi-unicondylar knee prosthesis functional assessment utilizing force-control wear testing

Author:

Spinelli M1,Affatato S1,Harman M K1,DesJardins J D2

Affiliation:

1. Laboratorio di Tecnologia Medica, Istituti Ortopedici Rizzoli, Bologna, Italy

2. Bioengineering Department, Clemson University, Clemson, USA

Abstract

Recent in vivo studies have identified variations in knee prosthesis function depending on prosthesis geometry, kinematic conditions, and the absence/presence of soft-tissue constraints after knee replacement surgery. In particular, unicondylar knee replacements (UKR) are highly sensitive to such variations. However, rigorous descriptions of UKR function through experimental simulation studies, performed under physiological force-controlled conditions, are lacking. The current study evaluated the long-term functional performance of a widely used fixed-bearing unicompartmental knee replacement, mounted in a bi-unicondylar configuration (Bi-UKR), utilizing a force-controlled knee simulator during a simulated (ISO 14243) walking cycle. The wear behaviour, the femoral—tibial kinematics, and the incurred damage scars were analysed. The wear rates for the medial and the lateral compartments were 10.27 ± 1.83 mg/million cycles and 4.49 ± 0.53 mg/million cycles, respectively. Although constant-input force-controlled loading conditions were maintained throughout the simulation, femoral—tibial contact point kinematics decreased by 65 to 68 per cent for average anterior/posterior travel and by 58 to 74 per cent for average medial/lateral travel with increasing cycling time up to 2 million cycles. There were no significant differences in damage area or damage extent between the medial and the lateral compartments. Focal damage scars representing the working region of the femoral component on the articular surface extended over a range of 16—21 mm in the anterior—posterior direction. Kinematics on the shear plane showed slight variations with increasing cycling time, and the platform exhibited medial pivoting over the entire test. These measures provide valuable experimental insight into the effect of the prosthesis design on wear, kinematics, and working area. These functional assessments of Bi-UKR under force-controlled knee joint wear simulation show that accumulated changes in the UKR articular conformity manifested as altered kinematics both for anterior/posterior translations and internal/external rotations.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3