Subcutaneous Implantable Cardioverter-Defibrillator Implantation in a Patient with a Left Ventricular Assist Device Already in Place

Author:

Gupta Ashwani,Subzposh Faiz,Hankins Shelley R.,Kutalek Steven P.

Abstract

A 56-year-old man with ischemic cardiomyopathy, a biventricular implantable cardioverter-defibrillator (ICD), and a left ventricular assist device (LVAD) developed a pocket hematoma and infection after an ICD generator change. The biventricular ICD was extracted, and the patient was given a full course of antibiotics. Because he had no indications for bradycardia pacing or biventricular pacing, he was implanted with a subcutaneous ICD under full anticoagulation. There was no interference in sensing or shock delivery from the ICD. The LVAD readings were unchanged during and after the procedure. The patient had an uneventful postoperative course, and both devices were functioning normally. To our knowledge, this is the first reported case of the implantation of a subcutaneous ICD in the presence of an LVAD. This report illustrates that both devices can be implanted successfully in the same patient. In addition, the subcutaneous ICD minimizes the risk of bloodstream infections, which can be fatal in patients who have life-supporting devices such as an LVAD.

Publisher

Texas Heart Institute Journal

Subject

Cardiology and Cardiovascular Medicine

Reference13 articles.

1. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to revise the ACC/AHA/NAPSE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons [published erratum appears in Circulation 2009;120(5):e34–5];Epstein;Circulation,2008

2. Risk of failure of transvenous implantable cardioverter-defibrillator leads;Borleffs;Circ Arrhythm Electrophysiol,2009

3. Predictors of short-term complications after implantable cardioverter-defibrillator replacement: results from the Ontario ICD Database;Krahn;Circ Arrhythm Electrophysiol,2011

4. An entirely subcutaneous implantable cardioverter-defibrillator;Bardy;N Engl J Med,2010

5. Left ventricular assist device in a patient with a concomitant subcutaneous implantable cardioverter defibrillator;Saeed;Circ Arrhythm Electrophysiol,2013

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3