Abstract
This study aimed to evaluate the general performance of the preclinical PET scanner (LabPET Solo 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 protocol. The performance parameters include: (i) spatial resolution; (ii) scatter fraction and counts rate performance; (iii) sensitivity and (iv) image quality. For the evaluation specific simulators and a sealed radioactive source of Sodium-22 were used. The results revealed that the NEMA-based radial spatial resolution ranged from 1.79 mm at the center of the FOV to 2.49 mm at a radial offset of 25mm. Tangential resolution ranged from 1.95mm at the center of the FOV to 3.50mm at a radial offset of 25mm. Axial resolution ranged from 1.78mm at the center of the FOV to 5.42mm at a radial offset of 25mm. Iterative image reconstruction improved the radial spatial resolution to 0.93mm at the center of the FOV. For the mouse-sized phantom, the peak noise equivalent count rate (NECR) was 37.8kcps at 85MBq whereas the peak true count rate was 169.9kcps at 125MBq with a count loss of 7.8% for the maximum rate reached at 103.3MBq. The peak of sensitivity (S) and absolute sensitivity (SA) at the center of axial field-of-view (FOV) is 5.7cps.Bq-1 and 0.63%, respectively. The overall imaging capabilities of the scanner were assessed using the NEMA image-quality phantom. The image roughness was 9.5%. Spillover rates (SORs) in air and in water were 0.26 and 0.17 respectively. The recovery coefficients (RCs) were 0.12 for the 1-mm-diameter rod and 0.91 for the 5-mm-diameter rod. The results demonstrate that CDTN/CNENsmall animal PET scanner has adequate reliability and acquired images are compatible with international quality standards for the molecular imaging research.
Publisher
Sociedade Brasilieira de Protecao Radiologica - SBPR
Reference31 articles.
1. [1] YAO R.; LECOMTE R.; CRAWFORD E. Small-Animal PET: What is it, and why do we need it? Journal of Nuclear Medicine Technology. Vol. 40 nº3, pp.157-165. 2012. https://doi.org/10.2967/jnmt.111.098632
2.
3. [2] GONTIJO, R. M. G. Proposta de Programa de Garantia da Qualidade para Imagem Mole-cular Pré-Clínica. 2019. 275 f. Doctoral Thesis (Doutorado em Ciência e Tecnologia das Radi-ações, Minerais e Materiais) - Centro de Desenvolvimento da Tecnologia Nuclear, Comissão Nacional de Energia Nuclear, Belo Horizonte, In Portuguese, 2019.
4.
5. [3] NEMA - National Electrical Manufacturers Association. Performance Measurements of Small Animal Positron Emission Tomographs. Rosslyn VA; 2008 Standards Publication NU 4-2008.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献