Affiliation:
1. Beihang Hangzhou Innovation Institute
Abstract
The cell temperature working point optimization of the spin-exchange relaxation-free (SERF) co-magnetometer is studied theoretically and experimentally in this article. Based on the steady-state solution of the Bloch equations, the steady-state response model of the K-Rb-21Ne SERF co-magnetometer output signal with cell temperature is established in this paper. And combined with the model, a method to find the optimal working point of the cell temperature that incorporates the pump laser intensity is proposed. The scale factor of the co-magnetometer under different pump laser intensities and cell temperatures is obtained experimentally, and the long-term stability of the co-magnetometer at the different cell temperatures with corresponding pump laser intensities is measured. The results show that the bias instability of the co-magnetometer is reduced from 0.0311 deg/h to 0.0169 deg/h by obtaining the optimal working point of the cell temperature, which verifies the validity and accuracy of the theoretical derivation and the proposed method.
Funder
National Science Fund for Distinguished Young Scholars
National Natural Science Foundation of China
Subject
Atomic and Molecular Physics, and Optics
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献