Affiliation:
1. Johor Bahru Clinical School, Monash University Sunway Campus, Bukit Azah, Johor Bahru, Johor - Malaysia
2. Department of Intensive Care, Austin Hospital, Heidelberg, Melbourne - Australia
Abstract
Background and Aims Extracorporeal cytokine removal may be desirable. We sought to assess extracorporeal blood purification (EBP) techniques for cytokine removal in experimental animal studies. Methods We conducted a targeted, systematic search and identified 17 articles. We analyzed cytokine clearance, sieving coefficient (SC), ultrafiltrate (UF) concentration, and percentage removal. As this review concerns technical appraisal of EBP techniques, we made no attempts to appraise the methodology of the studies included. Results are in descriptive terms only. Results Applying predicted clearance for 80 kg human, high volume hemofiltration (HVHF) techniques and plasmafiltration (PF) showed the highest rates of cytokine removal. High cutoff (HCO)/HF and PF techniques showed modest ability to clear cytokines using low to medium flows. Standard hemofiltration had little efficacy. At higher flows, HCO/HF achieved clearances between 30 and 70 ml/min for IL-6 and IL-10. There was essentially no removal of tumor necrosis factor (TNF)-alpha outside of PF. Conclusions Experimental animal studies indicate that HVHF (especially with HCO filters) and plasmafiltration have the potential to achieve appreciable IL-6 and IL-10 clearances. However, only PF can remove TNF-alpha reliably.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献