Affiliation:
1. School of Mechanical and Aerospace Engineering, Queen's University, Belfast - UK
2. Outcomes Assessment Unit, Musgrave Park Hospital, Belfast - UK
Abstract
The ability to measure acetabular cup orientation accurately during total hip arthroplasty represents a significant challenge. The aim of this research was to develop and evaluate a novel low cost mechanical device for measuring operative acetabular inclination. Cup implantation was simulated in two trials using the novel device: firstly involving surgeons and engineers orientating acetabular cups with sawbone pelves at a range of inclination angles (20°-55° in 5° increments); secondly in a simulated intra-operative scenario with surgeons. Target angles were compared with achieved angles and deviations from desired angles were recorded. In addition, all participants orientated cups under the same conditions using two other techniques: freehand and with a propriatory Mechanical Alignment Guide. In the first trial, the mean errors (deviations) using freehand technique, the mechanical alignment guide and the new device were 5.2° +/- 4.3° (range 0.1–22.0), 3.6° +/- 3.9° (range 0.1°–33.6°) and 0.5° +/- 0.4° (range 0.0–1.9) respectively. In the second trial, the mean error for freehand technique, mechanical alignment guide and the new device were 6.2° +/- 4.2° (range 0.2–18.2), 3.8° +/- 3.3° (range 0.0–19.1) and 0.6° +/- 0.5° (range 0.0–1.8) respectively. The new device has the potential to allow the surgeon to choose and record operative inclination accurately during total hip arthroplasty in the lateral decubitus position.
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献