Clinical Image Quality and Sensitivity in an Organized Mammography Screening Program

Author:

Théberge Isabelle12,Guertin Marie-Hélène1,Vandal Nathalie1,Daigle Jean-Marc1,Dufresne Michel-Pierre3,Wadden Nancy4,Shumak Rene5,Samson Caroline67,Langlois André1,Larocque Isabelle1,Perron Linda128,Pelletier Éric1,Brisson Jacques129

Affiliation:

1. Institut national de santé publique du Québec, Quebec City, Québec, Canada

2. Département de médecine sociale et préventive, Faculté de Médecine, Université Laval, Quebec City, Québec, Canada

3. Département de Radiologie, Hôpital Maisonneuve-Rosemont, Montréal, Québec, Canada

4. Faculty of Medicine, Memorial University, St John's, Newfoundland, Canada

5. Department of Prevention and Cancer Control, Cancer Care Ontario, Toronto, Ontario, Canada

6. Hôpital du Sacré-Cœur de Montréal, Montréal, Québec, Canada

7. Département de radiologie, radio-oncologie et médecine nucléaire, Faculté de Médecine, Université de Montréal, Montréal, Québec, Canada

8. Département de santé publique et médecine préventive, CHU de Québec-Université Laval, Hôpital Saint-François d'Assise, Quebec City, Québec, Canada

9. Axe Oncologie, Centre de Recherche du CHU de Québec, Quebec City, Québec, Canada

Abstract

Purpose The study sought to examine the association between clinical image quality of mammograms and screening sensitivity. Methods Four radiologists evaluated the clinical image quality of 374 invasive screen-detected cancers and 356 invasive interval breast cancers for which quality evaluation of screening mammograms could be assessed from cancers diagnosed among participants in the Quebec Breast Cancer Screening Program in 2007. Quality evaluation was based on the Canadian Association of Radiologists accreditation criteria, which are similar to those of the American College of Radiology. The association between clinical quality and screening sensitivity was assessed by logistic regression. Adjusted sensitivity and adjusted sensitivity ratios were obtained through marginal standardization. No institutional review board approval was required. Results A proportion of 28% (206 of 730) of screening mammograms had lower overall quality for the majority of assessments. Positioning was the quality attribute that was the most frequently deficient. The 2-year screening sensitivity reached 68%. Sensitivity of screening was not statistically associated with the overall quality (ratio of 2-year sensitivity = 1.03; 95% confidence interval: 0.93-1.15) or with any quality attributes (positioning, exposure, compression, sharpness, artifacts, contrast). Results were similar for the 1-year sensitivity. Conclusions Although not all mammograms in the Quebec screening program met the optimum quality required by the Canadian Association of Radiologists or American College of Radiology accreditation, the screening mammograms produced in this population-based organized screening program reached a high enough level of quality so that the remaining variation in quality is too little to impair screening sensitivity.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine

Reference25 articles.

1. PerryN. BroedersM. de WolfC. TörnbergS. HollandR. von KarsaL. European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis 2006 Oxford University Press Oxford, UK

2. Quality assurance for screening mammography: an international comparison

3. NoëlF. TremblayR. Technologue en radiologie Manuel de contrôle de la qualité en mammographie Vol. 1 2001 Ministère de la Santé et des Services Sociaux Québec, Canada

4. NoëlF. TremblayR. Physicien biomédical Manuel de contrôle de la qualité en mammographie Vol. 2 2006 Ministère de la Santé et des Services Sociaux Québec, Canada

5. Reasons for Failure of a Mammography Unit at Clinical Image Review in the American College of Radiology Mammography Accreditation Program