Affiliation:
1. Division of Combustion Physics, Lund University, Box 118, SE-221 00 Lund, Sweden (O.J., J.B., M.A.); and Tetra Pak Packaging Solutions, Ruben Rausings gata, SE-221 86, Lund, Sweden (U.L.)
Abstract
Photofragmentation laser-induced fluorescence (PF-LIF) is for the first time demonstrated to be a practical diagnostic tool for detection of hydrogen peroxide. Point measurements as well as two-dimensional (2D) measurements in free-flows, with nitrogen as bath gas, are reported. The present application of the PF-LIF technique involves one laser, emitting radiation of 266 nm wavelength, to dissociate hydrogen peroxide molecules into OH radicals, and another laser, emitting at 282.25 nm, to electronically excite OH, whose laser-induced fluorescence is detected. The measurement procedure is explained in detail and a suitable time separation between photolysis and excitation pulse is proposed to be on the order of a few hundred nanoseconds. With a separation time in that regime, recorded OH excitation scans were found to be thermal and the signal was close to maximum. The PF-LIF signal strength was shown to follow the same trend as the vapor pressure corresponding to the hydrogen peroxide liquid concentration. Thus, the PF-LIF signal appeared to increase linearly with hydrogen peroxide vapor-phase concentration. For 2D single shot measurements, a conservatively estimated value of the detection limit is 30 ppm. Experiments verified that for averaged point measurements the detection limit was well below 30 ppm.
Subject
Spectroscopy,Instrumentation
Cited by
27 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献