Affiliation:
1. Department of Chemistry, Saint Francis Xavier University, Antigonish, Nova Scotia B2G 1C0, Canada
Abstract
We have measured and compared ATR/FT-IR spectra of aqueous dispersions of colloidal silica, and aqueous suspensions of particulate silica, over ranges of bulk concentrations using CIRCLE® and Contact Sampler™ cells. Procedures are described for quantitative analyses of each silica type. Detection limits are approximately 0.001% for the particulate species. The spectra of particulates are distorted by refractive-index effects to low frequency of the main analytical bands around 1090 cm-1, but the distortions are proportional to concentration and do not interfere with the precision of analyses. We recommend either ATR cell, but the Contact Sampler™ is more convenient.
Subject
Spectroscopy,Instrumentation
Cited by
12 articles.