Affiliation:
1. Department of Applied Physics, Biomedical Technology Institute, University of Twente, P.O. Box 217, 7500 AE Enschede, the Netherlands
Abstract
A high-resolution confocal line-scan Raman microscope was developed for the study of biological samples such as cells and chromosomes. With the illumination of a line on the sample, all the spectral information of the line is recorded. The line-scan microscope was attained by the introduction of one scanning mirror, a slit, and two cylindrical lenses in a standard confocal Raman microscope. The resolution, for an ideal sample, is 0.5 μm in the lateral direction and 3.5 μm in the axial direction. Line-scan experiments were performed to demonstrate the applicability of the instrument for biological samples. Line-scan images of calcium phosphate bone implant coatings and of the banding pattern of polytene chromosomes are presented.
Subject
Spectroscopy,Instrumentation
Cited by
57 articles.