Affiliation:
1. Department of Chemistry, Florida State University, Tallahassee, Florida 32306-4390
Abstract
A procedure has been developed to permit recalculation of a Raman spectrum that has been taken on a source spectrometer to allow its quantitative use with data taken from a target system. The procedure is based on accurate characterization of both the source and target spectral abscissas by means of a neon spectrum that is recorded at the time the Raman measurements are made. Corrections are made for wavelength-dependent behavior of both systems, for differences in dispersion for the two systems, and for variation in the number, spacing, and size of the detector elements. An example is presented in which the fingerprint and CH regions of the Raman spectrum of cyclohexane are transferred from higher to lower resolution context.
Subject
Spectroscopy,Instrumentation
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献