Affiliation:
1. Institute of Hydrochemistry, Technical University of Munich, Marchioninistr. 17, D-81377 Munich, Germany
Abstract
Pulsed photoacoustic laser spectroscopy was used for depth-resolved analysis of artificial tissue models. The technique was applied to investigate the spatial resolution capabilities of a fiber-optical-coupled photoacoustic sensor head. The time-resolved measurements confirmed the theoretical predictions of a depth resolution of 0.1 mm. In an adapted skin model, a strongly absorbing target could be detected up to a layer depth of 14 mm. At a layer depth of 5 mm, a lateral resolution of 3.5 mm was achieved. Because of the depth-resolving capability, this method is well suited as a complementary approach for two-dimensional imaging techniques.
Subject
Spectroscopy,Instrumentation
Cited by
25 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献