Affiliation:
1. University of Arkansas, Department of Chemistry, Fayetteville, Arkansas 72701
Abstract
The design of a novel type of densitometer for evaluating photographic plates from gas electron diffraction is described. A commercially available linear diode array has been interfaced to a PDP-11/23 minicomputer via the DRV11 parallel interface. The diffraction pattern from a photographic plate is optically focused onto a solid state linear photodiode array. The position of each diode in the array provides the radial location information and the radial intensity of each diode is digitized with 14-bit resolution. Software development provides for control over the scan parameters for the array as well as for on-line manipulation of the collected data. The new instrument thus retains the advantages of digital data filtering described in a previous procedure, but allows for faster data recording since the total pattern is determined in one scanning measurement without the mechanical stepping process which is needed in single channel instruments to move the diffraction pattern through the measuring beam. Rapid acquisition of data with on-line transgeneration should prove to be valuable in the application of the new procedure as a real time data acquisition system for gas phase electron diffraction without photographic intensity recording.
Subject
Spectroscopy,Instrumentation
Reference19 articles.
1. Electron Diffraction as a Tool of Structural Chemistry
2. Processor-Controlled Microdensitometry
3. Nunn W. G., Dessy R. E., and Reynolds W. R., in Multichannel Image Detectors, ACS Symposium Series 102, Talmi Y., Ed. (American Chemical Society, Washington, DC, 1979), p. 136.
4. Electro-Optical Ion Detectors in Mass Spectrometry Simultaneous Monitoring of All Ions over Wide Mass Ranges
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献