Affiliation:
1. Department of Chemistry, University of Arkansas, Fayetteville, Arkansas 72701
Abstract
A new scheme of densitometric data collection is described which comprises a digital filtering technique and process control program. A 6502-based PET 2001/32K microcomputer and an Elscan model 2500 specular microdensitometer are interfaced through the standard 6522 VIA port chip. An Analog Devices ADC80/12 monolithic A/D converter is used to digitize the analog output of the light photometering system. Additionally, control of the optical scanner assembly is provided through processor-driven 4-phase TTL logic circuitry. The technique applies to any analog signal of interest, either for filtering purposes or for recording directly into memory for later waveform or transient analysis. The maximum data gathering rate is 40 kHz, limited by the A/D conversion time of 25 μs.
Subject
Spectroscopy,Instrumentation
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献