The cesium fountain clock as primary frequency standard is widely used in the areas, such as time-keeping system, satellite navigation, fundamental physics research, etc. The principle of operation of cesium fountain clock is introduced. The noise source and frequency shift term are ananlyzed. The major noise source influencing frequency stability are cold atom loading time, microwave phase noise related to Dick effect, and detection laser frequency noise. The major frequency bias influencing frequency uncertainty is blackbody radiation frequency shift,cold atom collision frequency shift,distributed cavity phase frequency shift and microwave leakage frequency shift.The key technique to achieve highperformance cesium fountain clock is sumerized. The application of cesium fountain clock is presented. The status of space cesium clock and future primary frequency standard of optical clock are shown.
Publisher
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Subject
General Physics and Astronomy
Reference49 articles.
1. Bureau International des Poids et Mesures The International System of Units (SI) (8th ed.) p113
2. Wang Z B, Zhao L, Wang S G, et al. 2014 Sci. China-Phys. Mech. Astron. 57 1788
3. Li H J, Tang G F, He F, et al. 2013 Sci. China-Phys. Mech. Astron. 56 1591
4. Ramsey N F 1950 Phys. Rev. 78 695
5. Dick G J 1987 Proc. 19th Precise Time and Time Interval (PTTI) Applications and Planning Metting Redondo Beach, USA, 1987 p133
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献