Intraoperative Blood Pressure Monitoring in Obese Patients

Author:

Schumann Roman,Meidert Agnes S.,Bonney Iwona,Koutentis Christos,Wesselink Wilbert,Kouz Karim,Saugel Bernd

Abstract

Background The optimal method for blood pressure monitoring in obese surgical patients remains unknown. Arterial catheters can cause potential complications, and noninvasive oscillometry provides only intermittent values. Finger cuff methods allow continuous noninvasive monitoring. The authors tested the hypothesis that the agreement between finger cuff and intraarterial measurements is better than the agreement between oscillometric and intraarterial measurements. Methods This prospective study compared intraarterial (reference method), finger cuff, and oscillometric (upper arm, forearm, and lower leg) blood pressure measurements in 90 obese patients having bariatric surgery using Bland–Altman analysis, four-quadrant plot and concordance analysis (to assess the ability of monitoring methods to follow blood pressure changes), and error grid analysis (to describe the clinical relevance of measurement differences). Results The difference (mean ± SD) between finger cuff and intraarterial measurements was −1 mmHg (± 11 mmHg) for mean arterial pressure, −7 mmHg (± 14 mmHg) for systolic blood pressure, and 0 mmHg (± 11 mmHg) for diastolic blood pressure. Concordance between changes in finger cuff and intraarterial measurements was 88% (mean arterial pressure), 85% (systolic blood pressure), and 81% (diastolic blood pressure). In error grid analysis comparing finger cuff and intraarterial measurements, the proportions of measurements in risk zones A to E were 77.1%, 21.6%, 0.9%, 0.4%, and 0.0% for mean arterial pressure, respectively, and 89.5%, 9.8%, 0.2%, 0.4%, and 0.2%, respectively, for systolic blood pressure. For mean arterial pressure and diastolic blood pressure, absolute agreement and trending agreement between finger cuff and intraarterial measurements were better than between oscillometric (at each of the three measurement sites) and intraarterial measurements. Forearm performed better than upper arm and lower leg monitoring with regard to absolute agreement and trending agreement with intraarterial monitoring. Conclusions The agreement between finger cuff and intraarterial measurements was better than the agreement between oscillometric and intraarterial measurements for mean arterial pressure and diastolic blood pressure in obese patients during surgery. Forearm oscillometry exhibits better measurement performance than upper arm or lower leg oscillometry. Editor’s Perspective What We Already Know about This Topic What This Article Tells Us That Is New

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3