Author:
Chapalain Xavier,Morvan Thomas,Gentric Jean-Christophe,Subileau Aurélie,Jacob Christophe,Cadic Anna,Caillard Anaïs,Huet Olivier
Abstract
Abstract
Background
In the perioperative setting, the most accurate way to continuously measure arterial blood pressure (ABP) is using an arterial catheter. Surrogate methods such as finger cuff have been developed to allow non-invasive measurements and are increasingly used, but need further evaluation. The aim of this study is to evaluate the accuracy and clinical concordance between two devices for the measurement of ABP during neuroradiological procedure.
Methods
This is a prospective, monocentric, observational study. All consecutive patients undergoing a neuroradiological procedure were eligible. Patients who needed arterial catheter for blood pressure measurement were included. During neuroradiological procedure, ABP (systolic, mean and diatolic blood pressure) was measured with two different technologies: radial artery catheter and Nexfin. Bland-Altman and error grid analyses were performed to evaluate the accuracy and clinical concordance between devices.
Results
From March 2022 to November 2022, we included 50 patients, mostly ASA 3 (60%) and required a cerebral embolization (94%) under general anaesthesia (96%). Error grid analysis showed that 99% of non-invasive ABP measures obtained with the Nexfin were located in the risk zone A or B. However, 65.7% of hypertension events and 41% of hypotensive events were respectively not detected by Nexfin. Compared to the artery catheter, a significant relationship was found for SAP (r2 = 0.78) and MAP (r2 = 0.80) with the Nexfin (p < 0.001). Bias and limits of agreement (LOA) were respectively 9.6 mmHg (− 15.6 to 34.8 mmHg) and − 0.8 mmHg (− 17.2 to 15.6 mmHg), for SAP and MAP.
Conclusions
Nexfin is not strictly interchangeable with artery catheter for ABP measuring. Further studies are needed to define its clinical use during neuroradiological procedure.
Trial registration
Clinicaltrials.gov, registration number: NCT05283824.
Publisher
Springer Science and Business Media LLC