Intrapatient Reproducibility of the BISxp® Monitor

Author:

Niedhart Dagmar J.1,Kaiser Heiko A.1,Jacobsohn Eric2,Hantler Charles B.3,Evers Alex S.4,Avidan Michael S.5

Affiliation:

1. Research Fellow.

2. Associate Professor.

3. Professor.

4. Professor and Chair, Department of Anesthesiology, Henry E Mallinckrodt Professor of Anesthesiology, Professor of Internal Medicine and Molecular Biology and Pharmacology, Washington University School of Medicine.

5. Assistant Professor, Department of Anesthesiology, Washington University School of Medicine.

Abstract

Background The Bispectral Index (BIS) reportedly reflects anesthetic depth. It is recommended that anesthetic agents should be titrated to maintain the BIS between 40 and 60 arbitrary BIS units during anesthesia. For anesthesia providers to follow this recommendation, the monitor should be predictably affected by different anesthetic agents and have good interpatient and intrapatient reproducibility. The authors hypothesized that when two BISxp devices (Aspect Medical Systems, Newton, MA) are placed concurrently on the same patient, their readings are concordant throughout the anesthetic period. Methods Simultaneous BIS recordings from two BISxp monitors were obtained during anesthesia at 5-s intervals from 12 participants. Results In total 22,860 concurrent paired BIS readings were obtained. For 10.7% of the time, there were sustained periods of 30 s or greater where the readings suggested a different depth of anesthesia. For 6% of the time, there were sustained periods of 30 s or greater where the readings differed by 10 or more arbitrary BIS units. The regression coefficient (R) for the two devices was 0.65 (range, 0.35-0.92). There was zero bias between the devices, and the 95% limits of agreement ranged between -18 and +17. Conclusion A conflicting anesthetic management was suggested by the simultaneous BIS readings 10.7% of the time. These results suggest that BISxp does not always provide a reproducible single number. Anesthesia providers should not rely exclusively on the BIS reading when assessing depth of anesthesia.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference20 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3