Busulfan Interlaboratory Proficiency Testing Program Revealed Worldwide Errors in Drug Quantitation and Dose Recommendations

Author:

Kweekel Dina M.12ORCID,McCune Jeannine S.3,Punt Arjen M.45,van Luin Matthijs26,Franssen Eric J.F.27

Affiliation:

1. Department of Clinical Pharmacy and Toxicology, Leiden University Medical Center, Leiden, The Netherlands;

2. Drug Analysis and Toxicology Division (KKGT) of the Dutch Foundation for Quality Assessment in Medical Laboratories (SKML), Utrecht, The Netherlands;

3. Department of Hematologic Malignancies Translational Sciences and Department of Hematologic Malignancies & HCT, Beckman Research Institute of City of Hope, Duarte, California;

4. Department of Clinical Pharmacy, Division of Laboratory Medicine and Pharmacy, University Medical Center Utrecht, Utrecht University, Utrecht, The Netherlands;

5. Central Diagnostic Laboratory, University Medical Center Utrecht, Utrecht University, Utrecht, The Netherlands;

6. Department of Clinical Pharmacology, University Medical Center Utrecht, Utrecht, The Netherlands; and

7. Department of Clinical Pharmacy OLVG, Amsterdam, The Netherlands.

Abstract

Background: The clinical outcomes of busulfan-based conditioning regimens for hematopoietic cell transplantation (HCT) have been improved by personalizing the doses to target narrow busulfan plasma exposure. An interlaboratory proficiency test program for the quantitation, pharmacokinetic modeling, and busulfan dosing in plasma was developed. Previous proficiency rounds (ie, the first 2) found that 67%–85% and 71%–88% of the dose recommendations were inaccurate, respectively. Methods: A proficiency test scheme was developed by the Dutch Foundation for Quality Assessment in Medical Laboratories (SKML) and consisted of 2 rounds per year, with each round containing 2 busulfan samples. In this study, 5 subsequent proficiency tests were evaluated. In each round, the participating laboratories reported their results for 2 proficiency samples (ie, low and high busulfan concentrations) and a theoretical case assessing their pharmacokinetic modeling and dose recommendations. Descriptive statistics were performed, with ±15% for busulfan concentrations and ±10% for busulfan plasma exposure. The dose recommendations were deemed accurate. Results: Since January 2020, 41 laboratories have participated in at least 1 round of this proficiency test. Over the 5 rounds, an average of 78% of the busulfan concentrations were accurate. Area under the concentration–time curve calculations were accurate in 75%–80% of the cases, whereas only 60%–69% of the dose recommendations were accurate. Compared with the first 2 proficiency test rounds (PMID 33675302, October, 2021), the busulfan quantitation results were similar, but the dose recommendations worsened. Some laboratories repeatedly submit results that deviated by more than 15% from the reference values. Conclusions: The proficiency test showed persistent inaccuracies in busulfan quantitation, pharmacokinetic modeling, and dose recommendations. Additional educational efforts have yet to be implemented; regulatory efforts seem to be needed. The use of specialized busulfan pharmacokinetic laboratories or a sufficient performance in busulfan proficiency tests should be required for HCT centers that prescribe busulfan.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Pharmacology (medical),Pharmacology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3