A New 35-mm Short Intra-aortic Balloon Catheter: A Suitable Option also for Non-Small-Sized Patients?

Author:

Gelsomino Sandro12,Lozekoot Pieter Willem1,Lorusso Roberto1,de Jong Monique Maria1,Parise Orlando1,Matteucci Francesco1,Lucà Fabiana1,Kumar Narendra1,Dehkord Eshan Sadeghian1,Romano Salvatore Mario2,Gensini Gian Franco2,La Meir Mark13,Maessen Josephus Gregorius1

Affiliation:

1. Maastricht University Medical Center, Maastricht, the Netherlands, Florence, Italy

2. Careggi Hospital, Florence, Italy

3. Department of Cardiothoracic Surgery, University Hospital, Brussels, Belgium.

Abstract

Objective Visceral ischemia can be a potentially life-threatening complication of intra-aortic balloon pump (IABP) support. A shorter IABP catheter might lead to a reduction of visceral complications. In this animal study, we evaluate the effects of a 35-mL short catheter in comparison with a 40-mL standard-sized catheter. Methods Eighteen healthy swine underwent 120-minute ligation of the left anterior descending coronary artery followed by 6 hours of reperfusion being supported by either a short IABP catheter (short group) (n = 6) or a long IABP catheter (long group) (n = 6) or with no assistance (controls) (n = 6). Hemodynamics, visceral and coronary flows, as well as biochemical markers were evaluated throughout the different phases of the protocol. Results Mesenteric flows increased significantly at reperfusion (P < 0.001 both) remaining constant afterward (all, P > 0.05) in the short group, while remaining significantly lower in the long group at the start of reperfusion, remaining constantly lower than the short group and controls (P < 0.001 vs short, P < 0.003 vs controls). In both long and short groups, catheters improved renal flows at reperfusion (P < 0.001 both) without any further variation (P < 0.05). In the short group, the flows were higher during the whole of reperfusion (all, P < 0.05). Intra-aortic balloon pump support improved hemodynamic indices and coronary blood flows during reperfusion to a similar extent in both the small and the long group (P > 0.05). Conclusions The short IABP catheter proved to be as effective as the standard-sized catheter in supporting hemodynamics and coronary circulation. Furthermore, it even improves visceral flows in comparison with conventional IABP catheters.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,General Medicine,Surgery,Pulmonary and Respiratory Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3