Author:
Sunkaraneni V.S.,Banerjee A.,Gray R.F.
Abstract
Cochlear implants have transformed the treatment of sensorineural hearing loss. They have few major complications. The authors describe the case of a man fitted with a cochlear implant who suffered a postoperative subdural haematoma. The haematoma is thought to have been caused by bleeding from emissary veins opened by the drill passages used to anchor the sutures for the receiver/stimulator. The authors have abandoned tie down sutures in cochlear implants, preferring an appropriately deep well with squared-off rims, which would secure the implant in place. They have had no further complications of this nature.