Accuracy and clinical implications of seven 25-hydroxyvitamin D methods compared with liquid chromatography–tandem mass spectrometry as a reference

Author:

Roth Heinz Jürgen1,Schmidt-Gayk Heinrich1,Weber Holger2,Niederau Christoph3

Affiliation:

1. Limbach Laboratory, Department of Endocrinology and Oncology, Im Breitspiel 15, 69126 Heidelberg, Germany

2. Labor Prof G Enders und Partner, Rosenbergstraße 85, 70193 Stuttgart, Germany

3. Labor Dr Niederau, Leopoldstr 10, 44147 Dortmund, Germany

Abstract

Background The most reliable assessment of vitamin D status is measurement of plasma 25-hydroxyvitamin D (25[OH]D) concentration. High variability in 25(OH)D measurements due to utilized test and assay technologies and the lack of standardization against reference materials and reference method often confounds proper assessment of vitamin D status. Methods We evaluated the accuracy of six routinely available methodologies: high-performance liquid chromatography (HPLC), the IDS-radioimmunoassay (IDS-RIA) and enzyme immunoassay (IDS-EIA), the Nichols Advantage automated protein-binding assay (Advantage), two versions of the DiaSorin automated immunoassay (Liaison 1 and Liaison 2) – and one prototype automated immunoassay (Elecsys) for assessment of the 25(OH)D3 status in a cohort of 300 randomly selected patients' samples compared with the reference method liquid chromatography-tandem mass spectrometry (LC-MS/MS). Results Passing-Bablok regression analysis demonstrated a slope for each method compared with LC-MS/MS that varied from 0.62 (IDS-EIA) to 1.0 (HPLC). The Advantage and the Liaison 1 showed significant deviation from linearity with highly variable individual results vs. the LC-MS/MS. Difference plots revealed a considerable persistent proportional bias for the IDS-RIA and IDS-EIA. All evaluated methods except HPLC demonstrated a more or less considerable deviation of individual 25(OH)D3 values compared with LC-MS/MS defined target concentrations. Conclusions Standardization of methods for the quantification of 25(OH)D on a human-based sample panel by means of LC-MS/MS would help to reduce the inter-method variability with respect to accuracy existing in 25(OH)D measurement considerably. However, there will still remain differences in the accuracy of methods utilizing sample purification before final quantification or immunological reaction when compared with those methods without separate sample purification.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3