Detection of Hearing Loss Using 2f2-f1 and 2f1-f2 Distortion-Product Otoacoustic Emissions

Author:

Fitzgerald Tracy S.1,Prieve Beth A.1

Affiliation:

1. Syracuse University, Syracuse, NY

Abstract

Although many distortion-product otoacoustic emissions (DPOAEs) may be measured in the ear canal in response to 2 pure tone stimuli, the majority of clinical studies have focused exclusively on the DPOAE at the frequency 2f1-f2. This study investigated another DPOAE, 2f2-f1, in an attempt to determine the following: (a) the optimal stimulus parameters for its clinical measurement and (b) its utility in differentiating between normal-hearing and hearing-impaired ears at low-to-mid frequencies (≤2000 Hz) when measured either alone or in conjunction with the 2f1-f2 DPOAE. Two experiments were conducted. In Experiment 1, the effects of primary level, level separation, and frequency separation (f2/f1) on 2f2-f1 DPOAE level were evaluated in normal-hearing ears for low-to-mid f2 frequencies (700–2000 Hz). Moderately high-level primaries (60–70 dB SPL) presented at equal levels or with f2 slightly higher than f1 produced the highest 2f2-f1 DPOAE levels. When the f2/f1 ratio that produced the highest 2f2-f1 DPOAE levels was examined across participants, the mean optimal f2/f1 ratio across f2 frequencies and primary level separations was 1.08. In Experiment 2, the accuracy with which DPOAE level or signal-to-noise ratio identified hearing status at the f2 frequency as normal or impaired was evaluated using clinical decision analysis. The 2f2-f1 and 2f1-f2 DPOAEs were measured from both normal-hearing and hearing-impaired ears using 2 sets of stimulus parameters: (a) the traditional parameters for measuring the 2f1-f2 DPOAE (f2/f1 = 1.22; L1, L2 = 65, 55 dB SPL) and (b) the new parameters that were deemed optimal for the 2f2-f1 DPOAE in Experiment 1 (f2/f1 = 1.073, L1 and L2 = 65 dB SPL). Identification of hearing status using 2f2-f1 DPOAE level and signal-to-noise ratio was more accurate when the new stimulus parameters were used compared with the results achieved when the 2f2-f1 DPOAE was recorded using the traditional parameters. However, identification of hearing status was less accurate for the 2f2-f1 DPOAE measured using the new parameters than for the 2f1-f2 DPOAE measured using the traditional parameters. No statistically significant improvements in test performance were achieved when the information from the 2 DPOAEs was combined, either by summing the DPOAE levels or by using logistic regression analysis.

Publisher

American Speech Language Hearing Association

Subject

Speech and Hearing,Linguistics and Language,Language and Linguistics

Reference76 articles.

1. Distortion product otoacoustic emission (2f1-f2) amplitude as a function of f2/f1 frequency ratio and primary tone level separation in human adults and neonates;Abdala C.;Journal of the Acoustical Society of America,1996

2. Micromechanical models of the cochlea;Allen J. B.;Physics Today,1992

3. Guidelines for screening for hearing impairment and middle ear disorders;ASHA,1990

4. Pure-tone threshold estimation from extrapolated distortion product otoacoustic emission I/O-functions in normal and cochlear hearing loss ears;Boege P.;Journal of the Acoustical Society of America,2002

5. Two sources of acoustic distortion products from the human cochlea;Brown A. M.;Journal of the Acoustical Society of America,1996

Cited by 11 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献