Evaluation of the Characteristics of Photon Radiation Fields of Medical Linear Electron Accelerators of Different Manufacturers

Author:

Taraev Aleksandr1,Zaharadniuk Aliaksei12,Bogdan Mihail1

Affiliation:

1. Scientific Production Unitary Enterprise «ATOMTEX»

2. Belarusian State University

Abstract

The article presents an assessment of the characteristics of pulsed photon radiation fields generated by medical linear electron accelerators (or CLINAC for short), namely: Varian Clinac iX, Elekta Infinity, Varian VitalBeam. Their technical characteristics, namely: average radiation energy, pulse duration and repetition rate, ambient dose equivalent rate in a single pulse are studied from the point of view of using a CLINAC as a source of pulsed photon radiation field suitable for calibration of dosimetric equipment. The average photon radiation energy is calculated using Monte Carlo modeling. Measurement of the ambient dose equivalent rate generated by the CLINAC at a given point is performed by the reference dosimeter DKS-AT5350/1. Determination of single pulse duration and pulse repetition rate is performed by direct measurements using the pulsed photon radiation detection unit BDKG-206.

Publisher

Scientific Production Company Doza, Ltd

Subject

General Engineering

Reference15 articles.

1. Загороднюк А.А., Тараев А.Ю., Лазаренко С.В., Комар Д.И. Влияние свинцового фильтра на среднюю энергию фотонного излучения медицинского линейного ускорителя электронов//АНРИ. 2023. №№2(113). С. 13-24., Zagorodnyuk A.A., Taraev A.Yu., Lazarenko S.V., Komar D.I. Vliyanie svincovogo fil'tra na srednyuyu energiyu fotonnogo izlucheniya medicinskogo lineynogo uskoritelya elektronov//ANRI. 2023. №№2(113). S. 13-24.

2. IEC TS 62743:2012. Radiation protection instrumentation – Electronic counting dosemeters for pulsed fields of ionizing radiation., IEC TS 62743:2012. Radiation protection instrumentation – Electronic counting dosemeters for pulsed fields of ionizing radiation.

3. U. Ankerhold et. al., «Deficiencies of active electronic radiation protection dosemeters in pulsed fields», Rad. Prot. Dosimetry, vol. 135, no. 3, 2009., U. Ankerhold et. al., «Deficiencies of active electronic radiation protection dosemeters in pulsed fields», Rad. Prot. Dosimetry, vol. 135, no. 3, 2009.

4. Мартынюк Ю.Н., Нурлыбаев К., Ревков А.А. Дозиметрия импульсного излучения//АНРИ. 2018. №№1(92). С. 2-11., Martynyuk Yu.N., Nurlybaev K., Revkov A.A. Dozimetriya impul'snogo izlucheniya//ANRI. 2018. №№1(92). S. 2-11.

5. Дозиметры рентгеновского и гамма-излучения ДКС АТ-1121, ДКС АТ-1123. URL: https://atomtex. com/ru/dozimetry-rentgenovskogo-i-gamma-izlucheniya-dks-at1121-dks-at1123 (дата обращения: 20.08.2023)., Dozimetry rentgenovskogo i gamma-izlucheniya DKS AT-1121, DKS AT-1123. URL: https://atomtex. com/ru/dozimetry-rentgenovskogo-i-gamma-izlucheniya-dks-at1121-dks-at1123 (data obrascheniya: 20.08.2023).