Author:
Muzamil Mustafa F,Shafii N S,Abdul Kadir M R,Che Aziz M N,Osman K,Dillon J,Darlis N,Md Khudzari A Z
Abstract
Abstract
Spiral flow-inducing cannula has been shown in previous research to exhibit a considerable effect on flow hemodynamic. However, there is still room for improvement. In this study, several design variations were tested to determine which variants were the best in terms of flow reduction. Computational Fluid Dynamics (CFD) software was used to simulate flow within a spiral flow-inducing cannula with several variations from chamber width and angle differences. The variants were compared against each other by using several flow parameters and a selection method was employed to determine which model was the best. It was found that a variant that has the widest chamber (14 mm) and biggest angle opening (70°) from the chamber to the cannula tube was the best in several parameters, and as such was chosen as the best variant. When compared with the standard straight cannula, the reduction in flow output was recorded to be 30% which is deemed significantly. In conclusion, spiral flow cannula recorded better hemodynamic effects with lower outflow velocity and wall shear stress value.