Author:
Roque Gerardo,Schütz Michael K.,Fey Julian,Useche Juan S.,Fiederle Michael,Procz Simon
Abstract
Abstract
The presented work describes the capabilities of a
state-of-the-art spectroscopic photon-counting hybrid pixel detector
for use in non-destructive testing, as a contaminant detector with
potential uses in the food and pharmaceutical industries. A
pharmaceutical hard capsule containing vitamin powder, contaminated
with Steel (ST) and Tungsten (W) was prepared under controlled
conditions. Both contaminants are distinguished from the powder by
combining sub-pixel imaging with spectroscopic imaging analysis,
made possible by the use of a 2 mm CdZnTe sensor coupled to a
Timepix3 ASIC, operated in Time-over-Threshold and Time-of-Arrival
mode. The sub-pixelization technique achieves a spatial resolution
of 18.3 μm, with a maximum photon flux of
2.5× 106 photons/s· cm2. For the spectral
analysis technique implemented, each energy bin image is
energy-weighted, with weights optimally defined by considering the
bin contrast and noise behavior, and combined to form an
integrated-spectrum enhanced image with increased contaminant
visibility. Contrast-to-Noise Ratio (CNR) for each contaminant type
was used for performance assessment of the approach. A CNR increase
of 106% above the reference unweighted-spectrum image was achieved
for the ST contaminant, while the denser W contaminant showed a CNR
increase of 80%. This means that the presented spectroscopic
imaging technique is feasible for implementation in material
visibility enhancement.