Abstract
Abstract
In particle therapy of lung tumors, modulating effects on the particle beam may occur due to the microscopic structure of the lung tissue. These effects are caused by the heterogeneous nature of the lung tissue and cannot be completely taken into account during treatment planning, because these micro structures are too small to be fully resolved in the planning CT. In several publications, a new material parameter called modulation power (P
mod
) was introduced to characterize the effect. For various artificial lung surrogates, this parameter was measured and published by other groups and ranges up to approximately 1000 μm. Studies investigating the influence of the modulation power on the dose distribution during irradiation are using this parameter in the rang of 100–800 μm. More precise measurements for P
mod
on real lung tissue have not yet been published. In this work, the modulation power of real lung tissue was measured using porcine lungs in order to produce more reliable data of P
mod
for real lung tissue. For this purpose, ex-vivo porcine lungs were frozen in a ventilated state and measurements in a carbon ion-beam were performed. Due to the way the lungs were prepared and transferred to a solid state, the lung structures that modulate the beam could also be examined in detail using micro CT imaging. An optimization of the established methods of measuring the modulation power, which takes better account of the typical structures within lung tissue, was developed as well.
Funder
Bundesministerium für Bildung und Forschung
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献