Accelerated 3D image reconstruction with a morphological pyramid and noise-power convergence criterion

Author:

Sisniega AORCID,Stayman J W,Capostagno S,Weiss C R,Ehtiati T,Siewerdsen J HORCID

Abstract

Abstract Model-based iterative reconstruction (MBIR) for cone-beam CT (CBCT) offers better noise-resolution tradeoff and image quality than analytical methods for acquisition protocols with low x-ray dose or limited data, but with increased computational burden that poses a drawback to routine application in clinical scenarios. This work develops a comprehensive framework for acceleration of MBIR in the form of penalized weighted least squares optimized with ordered subsets separable quadratic surrogates. The optimization was scheduled on a set of stages forming a morphological pyramid varying in voxel size. Transition between stages was controlled with a convergence criterion based on the deviation between the mid-band noise power spectrum (NPS) measured on a homogeneous region of the evolving reconstruction and that expected for the converged image, computed with an analytical model that used projection data and the reconstruction parameters. A stochastic backprojector was developed by introducing a random perturbation to the sampling position of each voxel for each ray in the reconstruction within a voxel-based backprojector, breaking the deterministic pattern of sampling artifacts when combined with an unmatched Siddon forward projector. This fast, forward and backprojector pair were included into a multi-resolution reconstruction strategy to provide support for objects partially outside of the field of view. Acceleration from ordered subsets was combined with momentum accumulation stabilized with an adaptive technique that automatically resets the accumulated momentum when it diverges noticeably from the current iteration update. The framework was evaluated with CBCT data of a realistic abdomen phantom acquired on an imaging x-ray bench and with clinical CBCT data from an angiography robotic C-arm (Artis Zeego, Siemens Healthineers, Forchheim, Germany) acquired during a liver embolization procedure. Image fidelity was assessed with the structural similarity index (SSIM) computed with a converged reconstruction. The accelerated framework provided accurate reconstructions in 60 s (SSIM = 0.97) and as little as 27 s (SSIM = 0.94) for soft-tissue evaluation. The use of simple forward and backprojectors resulted in 9.3× acceleration. Accumulation of momentum provided extra ∼3.5× acceleration with stable convergence for 6–30 subsets. The NPS-driven morphological pyramid resulted in initial faster convergence, achieving similar SSIM with 1.5× lower runtime than the single-stage optimization. Acceleration of MBIR to provide reconstruction time compatible with clinical applications is feasible via architectures that integrate algorithmic acceleration with approaches to provide stable convergence, and optimization schedules that maximize convergence speed.

Funder

Siemens Healthineers

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3