Validation of an MR-based multimodal method for molecular composition and proton stopping power ratio determination using ex vivo animal tissues and tissue-mimicking phantoms

Author:

Marants Raanan,Tattenberg Sebastian,Scholey Jessica,Kaza Evangelia,Miao Xin,Benkert Thomas,Magneson Olivia,Fischer Jade,Vinas Luciano,Niepel Katharina,Bortfeld Thomas,Landry GuillaumeORCID,Parodi KatiaORCID,Verburg Joost,Sudhyadhom AtcharORCID

Abstract

Abstract Objective. Range uncertainty in proton therapy is an important factor limiting clinical effectiveness. Magnetic resonance imaging (MRI) can measure voxel-wise molecular composition and, when combined with kilovoltage CT (kVCT), accurately determine mean ionization potential (I m ), electron density, and stopping power ratio (SPR). We aimed to develop a novel MR-based multimodal method to accurately determine SPR and molecular compositions. This method was evaluated in tissue-mimicking and ex vivo porcine phantoms, and in a brain radiotherapy patient. Approach. Four tissue-mimicking phantoms with known compositions, two porcine tissue phantoms, and a brain cancer patient were imaged with kVCT and MRI. Three imaging-based values were determined: SPRCM (CT-based Multimodal), SPRMM (MR-based Multimodal), and SPRstoich (stoichiometric calibration). MRI was used to determine two tissue-specific quantities of the Bethe Bloch equation (I m , electron density) to compute SPRCM and SPRMM. Imaging-based SPRs were compared to measurements for phantoms in a proton beam using a multilayer ionization chamber (SPRMLIC). Main results. Root mean square errors relative to SPRMLIC were 0.0104(0.86%), 0.0046(0.45%), and 0.0142(1.31%) for SPRCM, SPRMM, and SPRstoich, respectively. The largest errors were in bony phantoms, while soft tissue and porcine tissue phantoms had <1% errors across all SPR values. Relative to known physical molecular compositions, imaging-determined compositions differed by approximately ≤10%. In the brain case, the largest differences between SPRstoich and SPRMM were in bone and high lipids/fat tissue. The magnitudes and trends of these differences matched phantom results. Significance. Our MR-based multimodal method determined molecular compositions and SPR in various tissue-mimicking phantoms with high accuracy, as confirmed with proton beam measurements. This method also revealed significant SPR differences compared to stoichiometric kVCT-only calculation in a clinical case, with the largest differences in bone. These findings support that including MRI in proton therapy treatment planning can improve the accuracy of calculated SPR values and reduce range uncertainties.

Funder

Massachusetts General Hospital

National Cancer Institute

National Institute of Biomedical Imaging and Bioengineering

Deutsche Forschungsgemeinschaft

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献